Fastq to Annotated Somatic Variants (Tumor Normal) with Coverage Analysis

The Fastq to Annotated Somatic Variants (Tumor Normal) with Coverage Analysis template workflow:

The workflow can only be used with targeted data.

The runtime of this workflow is significantly longer than the runtime of Fast to Annotated Somatic Variants (Tumor Normal) (Fastq to Annotated Somatic Variants), because a read mapping track is saved.

Fastq to Annotated Somatic Variants (Tumor Normal) with Coverage Analysis can be found at:

        Template Workflows | LightSpeed Workflows (Image lightspeed_wf_folder_open_16_n_p) | Fastq to Annotated Somatic Variants (Tumor Normal) with Coverage Analysis (Image annotate_variants_ls_16_n_p)

If you are connected to a CLC Server via your Workbench, you will be asked where you would like to run the analysis. We recommend that you run the analysis on a CLC Server when possible.

In the first wizard step, select the target regions (figure 4.22).

Image fatnvca_select_target_regions
Figure 4.22: Select the target regions.

Next, select a Reference Data Set (figure 4.23). If you have not downloaded the Reference Data Set yet, the dialog will suggest the relevant data set and offer the opportunity to download it using the Download to Workbench button.

If none of the available reference data sets are appropriate, custom reference data sets can be created, see http://resources.qiagenbioinformatics.com/manuals/clcgenomicsworkbench/current/index.php?manual=Custom_Sets.html.

Image fatnvca_select_ref_data
Figure 4.23: Select a reference data set.

In the LightSpeed Fastq to Somatic Variants (Tumor Normal) wizard step (figure 4.24) you have the following options:

Image lis_workflow_tn_cov_step3
Figure 4.24: Select fastq files.

In the next two wizard steps, define the threshold for minimum coverage for the normal (figure 4.25) and for the tumor (figure 4.26), respectively. These thresholds are important because they are used in the quality control section of the sample report. In the later wizard step for Create Sample Report, you will be able to adjust the percent bases in the target regions that should meet the thresholds.

Image fatnvca_set_coverage_threshold_normal
Figure 4.25: Set the coverage threshold for the normal sample. This threshold is used in the quality control section of the sample report.

Image fatnvca_set_coverage_threshold_tumor
Figure 4.26: Set the coverage threshold for the tumor sample. This threshold is used in the quality control section of the sample report.

In the Create Sample Report wizard step, select relevant summary items and specify thresholds for quality control (figure 4.27). Summary items, thresholds and an indication of whether specified thresholds were met, will be shown in the quality control section of the sample report. The default summary items are appropriate for many data sets, but may need to be adjusted.

To add more summary items, press Add..., choose the report type Lightspeed fastq to somatic variants tumor normal and select summary items as appropriate.

There are two instances of QC for Targeted Sequencing in the workflow. One for the normal read mapping and one for the tumor read mapping. To be able to add information from both the normal and the tumor reports from QC for Targeted Sequencing to the sample report, the reports are retyped using the Modify Report Type tool. To select additional summary items for the two reports, the following steps are needed: After pressing Add..., select the QC for targeted sequencing report type. Then check "Apply to custom report type" and write either "QC for Targeted Sequencing (Normal)" or "QC for Targeted Sequencing (Tumor)" as relevant before selecting summary items.

Image fatnvca_create_sample_report
Figure 4.27: Specify summary items. These will be shown in the quality control section of the sample report.

In the final wizard step, choose to Save the results of the workflow and specify a location in the Navigation Area before clicking Finish.



Subsections