Type with Consensus Refinement parameters

To run the tool, go to:

        Tools | Microbial Genomics Module (Image mgm_folder_closed_flat_16_h_p) | Typing and Epidemiology (Image typing_epi_folder_closed_16_h_p) | Type with Consensus Refinement (Image typing_using_consensus_16_h_p)

In the first dialog, select the reads for analysis, and click on Next.

Click Next to select a reference database. The reference database is a sequence list in which each sequence must be annotated with a Segment attribute. The tool uses the Segment information to define which sequences will be considered together when evaluating which of them suits better the input data. The sequences in the reference may also be annotated with a Type attribute (e.g. A and B) and a Subtype attribute (e.g. H1 and H5). Sequence lists can be updated manually in the Table View or using a table file with Update Sequence Attributes in Lists.

A ready-to-use reference data set for typing Influenza A, B, C and D is available through the Reference Data Manager.

In the Mapping options dialog, specify settings for the read mapping (figure 8.10). The options are identical to those of the Map Reads to Reference tool, except for the "Non-specific match handling".

Image type_with_consensus_step_1
Figure 8.10: Select mapping options.

The Filters dialog holds the following options (figure 8.11):

Image type_with_consensus_step_2
Figure 8.11: Choose your output option before saving your results.

Note that all filters apply per sequence, except for the Minimum relative abundance and Maximum number of references to report filters which consider all sequences with the same Segment and Type (if present). In the Minimum relative abundance case, if four references for Influenza A segment 3 are available for example, the filter considers the relative abundance among these four sequences and ignores the abundance in the rest of the references. In the Maximum number of references to report case, if the value is set to three, for a virus with eight segments such as Influenza, up to 24 segments may be reported (up to three per segment).